Specifications 

1.6.1 F-MXandF-MXGframes 

  • Power supply
    Rated voltages and frequencies: 100 to 240 VAC 60/50 Hz Allowed voltage fluctuations: ± 10%
    Maximum power consumption: 80 VA
    Fuses (2): T2AH/250V
    Environmental conditions
    Operating temperature: +10 to +40 °C (50 to 104 °F)
    Storage and transport temp: –10 to +50 °C (14 to 122 °F)
    Relative humidity: 10 to 85 % noncondensing, in airway 0 to 100 % condensing Atmospheric pressure: 660 to 1060 hPa (500 to 800 mmHg) 
  • Mechanics
    Dimension: 330 mm × 220 mm × 300 mm (width × depth × height)
    Weight: <11.2 kg / <24.8 lbs (F-MXG with all options); <10.2 kg / <22.6 lbs (F-MX with all options)
  • LCD display
    Display size: 10.4 inch
    Display type: Active matrix color LCD display Resolution: 640 × 480
  • Battery
    Type: 12V 2.6AH, lead acid
    Back-up battery time: at least 15 minutes when fully charged
    Charging time: 5 hours (typical)
    The green battery charge status LED is On when the battery is fully charged, on the holding voltage. The LED flashes when the battery is being charged.
 

1.6.2 NIBP

Measurement range:
adult 25 to 260 mmHg (3.3 to 34.7 kPa) child 25 to 195 mmHg (3.3 to 26.0 kPa) infant 15 to 145 mmHg (2.0 to 19.3 kPa)
Pulse rate range accepted: 30 to 250 bpm
Typical measuring time: adults 23 seconds, infants 20 seconds

1.6.3 Temperature

Measurement range: 10 to 45°C (50 to 113°F)
Measurement accuracy: 25 to 45.0 °C ± 0.1 °C (77 to 113 °F ± 0.2 °F) 10to24.9°C±0.2°C(50to76.8°F±0.4°F)
Probe type: Compatible with Datex-Ohmeda temperature probes only

1.6.4 ECG

Waveform display (with 50 Hz power supply frequency): Monitoring filter: 0.5 to 30 Hz
ST filter: 0.05 to 30Hz
Diagnostic filter: 0.05 to 100 Hz

Waveform display (with 60 Hz power supply frequency): Monitoring filter: 0.5 to 40 H
ST filter: 0.05 to 40 Hz
Diagnostic filter: 0.05 to 100 Hz

Heart rate
Measurement range: 30 to 250 bpm Measurement accuracy: ± 5% or ± 5 Pacemaker pulse detection level: 2 to 500 mV Pacemaker pulse duration: 0.5 to 2 ms

Overview
Cardiocap/5 Technical Reference Manual

1.6.5 Impedancerespiration

Respiration range: 4 to 120 respirations/minute Accuracy: ± 5% or ± 5 bpm

1.6.6 Pulseoximetry,standard

Display update time: 5 seconds
Averaging time: adjustable
Plethysmographic waveform scaling: adjustable

SpO2
Calibration range: 50 to 100% Calibrated against functional saturation Measurement range: 40 to 100%
Measurement accuracy (% SpO2 ±1 SD): 80 to 100% ± 2 digits;
50 to 80% ± 3 digits;
Below 50% unspecified

NOTE: SpO2 measurement accuracy is based on deep hypoxia studies using Datex-Ohmeda FingerSat sensors on volunteered subjects. Arterial blood samples were analyzed by a Radiometer OSM CO-oximeter. Refer to the sensor instructions for specific SpO2 accuracy data.

Pulse rate
Measurement range: 30 to 250 bpm Measurement accuracy: ± 5% or ± 5 bpm

Default alarm limits
SpO2: high Off, low 90% Pulse rate: high 160, low 40 NOTE: Limits are adjustable.
Sensor emitter wavelength ranges
Red LED: 660 nm Infrared LED: 900 nm